EM TEST
CNI 501 / CNI 503
Manual of operation
V 3.32
15 / 41
4.4.
General
DC supply
as per tables in 4.2
Coupling modes Surge
IEC 61000-4-5
Line to Ground all combinations
coupling capacitor 9
F
Source impedance 12
Line to Line
all combinations
coupling capacitor 18
F
Source impedance 2
Output sockets for testing signal and data lines with CNV 504 and CNV 508
Coupling modes Burst
IEC 61000-4-4
Line(s) to ground all combinations
Output sockets for capacitive coupling clamp
Remote control
UCS 500;
all models
Burst and Surge
EFT 500,
all models
Burst
VCS 500M; all models
Surge
Note:
In case that all remote control input at the CNI 503 are available and the user has all above
mentioned generators available a simultaneous use of all generators is not allowed.
The operator has to decide either to use the UCS 500 or the EFT/VCS combination of generators.
Power supply
Mains
115 V / 230 V
Fuse
2 A slow blow all models
Dimension and weight
Housing
19” 6HU (models up to 32 A all models if not otherwise specified)
Weight
16 A
models 3-ph: Approx. 30 kg
32 A
models 3-ph: Approx. 45 kg
models 1-ph approx. 30 kg