EM TEST
CNI 501 / CNI 503
Manual of operation
V 3.32
9 / 41
2.2.
Rear side
Figure 2.2: Rear side CNI 503 (4kV version)
Figure 2.3: CNI 503 Rear side (> 4kV version)
1
EUT power L1, L2, L3
2
EUT power N
3
EUT power PE
4
Ground Reference
5
HV Input from generator
5
HV Input from generator
6
COM Input from generator
7
Power switch with fuse
8
Voltage selector 230V / 115V
9
CN control input from UCS generator
10
CN control input from EFT and or VCS generator
If the CNI 503 includes two or more CN interfaces for
UCS500NE
and
UCS500NV
, only one
generator is allowed to communicate to the CNI 503. In case that all remote control input
, for UCS
(9) and EFT/VCS (10),
are included in the CNI 503 the operator has to take care that either the
UCS or the EFT/VCS is controlling the CNI.
A simultaneous operation by more than one generators is not allowed.
The CN control input is available in
two different versions
.
- Interface to generators of the series
UCS 500xx
- Interface to generators of the series
EFT 500x
and or
VCS500x
The connectors have different pin configurations. Therefore each generator
must be plugged in as signed at the rear panel.