Maintenance
31
Take care for the cartridge body to have the smaller diameter up and the bigger
down.
4.6 Combustion tube replacing
DANGER
G0002
Mortal danger from electric shock
Exposed power contacts - High Voltage
-
An electric shock can cause injuries in the form of burns and
cardiac arrhythmia, respiratory arrest or cardiac arrest.
•
Disconnect the mains power plug before opening the
analyzer’s cabinet.
Fig. 7: Removing the combustion tube
•
Switch off the analyzer and disconnect the power plug.
•
Remove the cover panel (1) of the cabinet. Unscrew 4 screws (2) on
top.
•
Remove the tubes (3), (4) and (5) from the upper furnace closure
assembly (6/7).
•
Remove the 4 screws (8) which hold the assembly (6/7) on the
furnace. The 4 screws and nuts holding the disk (7) on the part (6)
should remain in place.
•
Lift the upper furnace closure. The assembly consisting of the parts
(6), (7), (9), (10) and (11) is then taken out of the furnace. Take care