Installation
16
Fig. 6: Window Configuration – Heater - Preheating
The value shows the chosen heating rate in °C/min. (The preheating function is
switched off by entering 0 in this place. Then, the heating occurs with maximum
speed after start of the UNI-software.)
3.
Save your entry with Enter.
4.
The preheating starts automatically with the next start of the software, if the
furnace temperature is lower than 100°C and the analyzer switch is at
position 2.
NOTICE
When the preheating function is activated (by any other value except 0) the
software automatically creates a shortcut at the startup menu of Windows. This
enables the start of the UNI-software with every start of the PC. The shortcut is
deleted automatically when the preheating is deactivated.
2.9.2 BIOS-configuration
In order to start the preheating before the work begins, the BIOS has to be
configured to start the PC automatically at the selected time. (The preheating starts
automatically by starting the UNI-software.)
1.
Switch on your PC and enter the BIOS by pressing the corresponding key
(e.g. F2). The normal booting process is interrupted and the BIOS menu
appears.
2.
Choose Power Management Setup
3.
Activate "Resume by Alarm" (enabled).
4.
Enter the required days and times when the heating up of the analyzer
should be started.
5.
Save your entries at Save/Exit with Yes.
6.
The booting process will be sustained.
Содержание CS-580A
Страница 2: ...2 Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...
Страница 3: ...3 ...
Страница 23: ...Analysis 23 Weight 350mg Set the minimum analysis time to 50s Set the comparator level to 20mV ...
Страница 38: ...Function description 38 Fig 14 Gas flow ...
Страница 48: ...Ordering numbers 48 6 6 TIC module Fig 17 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ...Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...