52
If the display is showing
‘O/RANGE’ when the
set
button is pressed, the
display will show
‘CANNOT SET’ to indicate that the probe cannot be
calibrated. The Squirrel retains its previous calibration setting. Select the
next pH channel with the select button, and calibrate it in the same way.
This must be done for all the channels that are required.
9.5 Calibrating the Slope
The second calibration point is the slope. The pH of the buffer solution
used in this function should be set in the STAT function, as described
above. Once that is set as desired, calibration of the slope can be
performed in this function. The probes must be rinsed, placed in the
specified buffer solution, and allowed time to settle. If the temperature
compensation is being measured, then the temperature probes must be
put in the same solution.
S L O P E 9 c h 9
9 . 3 1 p H
Once the probes have settled, the display should show
‘SLOPEn’ (where
‘n’ is 4, 9 or 10). If this does not match the solution that is being used,
then the slope value in the STAT function must be changed appropriately
(see above). If it reads
‘O/RANGE’, then the pH input is too far out of
range of the preset value for the Squirrel to be calibrated to it. Again, this
is typically caused by a pH probe that has expired. Once the value has
settled, press and hold the
set
button.
‘HOLD TO CHANGE’ will appear on
the display and then the logger will recalibrate itself on the current channel
to read the correct value (in this example, 9.00 pH): -
S L O P E 9 c h 9
9 . 0 0 p H
If the display is showing
‘O/RANGE’ when the
set
button is pressed, the
display will show
‘CANNOT SET’ to indicate that the probe cannot be
calibrated. The Squirrel retains its previous calibration setting. Select the
next pH channel with the select button, and calibrate it in the same way.
This must be done for all the channels that are required.
9.6 Current Consumption and Battery Life
Due to the sensitive nature of pH probes, the Squirrel has to provide a
constant power supply to the pH input circuitry when pH inputs are in use.
This power supply is switched on when one or more of the pH channels