![Eltek Squirrel 1000 Series Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/eltek/squirrel-1000-series/squirrel-1000-series_manual_2397252039.webp)
39
5.8 Wiring up a pulse count channel
Pulse count channels may be wired for a voltage input or a voltage-free
contact (see section 4.9.3 for details of voltages and mechanical contact
conditions). The positive connection should be made to the
‘+’ input and
the negative to the
‘-’ input. For a voltage-free contact input (e.g. relay),
the contact
‘debounce’ circuit must be connected. This involves linking the
‘-’ pin to the ‘s’ pin: -
s
s
sensor output
volt-free contact
Note: the
‘-’ connection is ground and should be connected to the cable
screen, if any.
5.8.1 When 2 pulse channels are fitted, the connector is a 4 way:
+ -
S
x
sensor output
+
-
X = the additional channel input and is a function of the logger type. It is
configured for 200 Hz max frequency input and has debounce voltage-
free contacts fitted internally.