10
The temperature range then increments to the next available range with
every press of the set button: -
C H 1 ° F
- 2 2 . 0 / 1 5 8 . 0
Leave the display showing your chosen range and press select to set up
another channel.
3.4.2 Voltage/current channel setup
These channels may be configured for voltage or current. There are
typically 18 voltage ranges and 2 current ranges. Voltage/current
channels have an additional range called the EU (Engineering Units)
range where the user can select specific engineering units (e.g. m/s) and
a max/min scale (e.g. 0/50 m/s). The appropriate voltage/current range is
then selected from the existing hardware ranges (e.g. 0/1 V). In the above
example, 0/50 m/s = 0/1V and the display in METER mode will show
readings in m/s. All data in this channel will be stored and downloaded
using m/s as the unit. In the above example, the CHANNEL SETUP
function will alternate between the following displays: -
C H 6 E U m / s
0 . 0 / 5 0 . 0
C H 6 E U V
0 . 0 0 0 / 1 . 0 0 0
The EU range may only be set up using Darca (please see price list or
contact your distributor for details). Whenever a voltage/current channel
has not been set up, the initial range is always the default EU range and is
displayed as follows: -
C H 6 E U %
0 . 0 / 1 0 0 . 0
C H 6 E U V
0 . 0 0 0 / 1 . 0 0 0
Press
select
to change channel. For example, select channel 5: -