Operation
45
6.3.6.2 Certification data log
The certification data log is activated by default, but can be deactivated as an option. The
parameters affected are then below the calibration lock.
Using the certification data log in accordance with PTB
-
A 50.7, some calibration law-relevant
parameters can also be changed when the calibration lock is closed. The requirements for
this are:
The administrator lock must be opened (see below).
At least three free entries must be available in the certification data log.
The parameters affected according to the factory settings are:
CP.I1
cp value for input 1,
MPer
measurement period
Md.K
K number mode
Clr.A
delete measurement archive
Md.I1
mode for input 1
QMax
maximum load (plausibility check in encoder mode)
For every change of a parameter under
CDL
with a closed calibration lock, a data row for the
value is entered before and after the change.
If the certification data log is full, it can be deleted when the calibration lock is open using the
ClCDL
command. (see chapter 5.3.1.13)
If the calibration lock is opened while the calibration logbook is open, it can only
be closed again after deleting the certification data log.
6.3.6.3 Administrator lock and customer lock
Administrator and customer lock are used to secure all calibration law data that are not
relevant, but that should not be changed without authorization.
The locks can be opened by entering a code (the "key") under
Cod.A
or
Cod.C
and closed by
entering "0" for
St.AL
or
St.CL
.
These values can be found under the path:
Admin
Device settings
Access
or
Serv
.
Device settings
Access
There, a time can also be set in minutes for each lock after which it closes automatically.
Содержание EK205
Страница 1: ...Operating manual Volume conversion device EK205 ...
Страница 6: ......
Страница 64: ...64 Appendix 9 2 Conformity declarations19 19 IECEx is not available for all device types ...
Страница 65: ...Appendix 65 9 3 Design test certifications 9 3 1 ATEX EC Type Examination Certificat ...
Страница 66: ...66 Appendix ...
Страница 67: ...Appendix 67 ...
Страница 68: ...68 Appendix 9 3 2 IECEx Certificate of Conformity20 20 Not available for all device types ...
Страница 69: ...Appendix 69 ...
Страница 70: ...70 Appendix ...
Страница 71: ...Appendix 71 ...