Installation

, connection and commissioning 

25 

5.1.3 Installation on a wall 

 

Using the wall mounting feet (see appendix), install the EK205 on the wall.  

 

 

 

 

Ill.: 6

 

1.  Using the four self-cutting screws 

provided on the rear of the EK205

install the wall mounting feet. The 
fastening feet can be fastened in 
different positions.  

2.

 

Drill holes in the wall

 

in accordance 

with the position of the installed 
installation feet.

 

Select the wall plugs 

suitable for the size of the screws and 
insert them in the borings of the well. 

3.

 

Use four

 

screws for fastening the 

EK205, e.g. 5 x 35 mm. 

 

5.1.4 Three-way cock

7

 

When installing the pressure sensor, a three-way cock is usually installed in order to carry out 
an inspection of the pressure sensor in built-in condition or for the replacement of deficient 
sensors without deactivating the overall gas line. The three-way cock available from Elster 
has the following structure: 

 

 

 

The pipeline from the pressure sensor to the meter must be laid in a falling 
manner so that, where applicable, water cannot damage the pressure sensor 
or influence the measurement precision. 

 

 

                                                      

7

 Not applicable when using EK205 as a Temperature Conversion Device. 

 

Ill.: 7 

 1 

:

 

From the meter: From the pressure 
connection of the gas meter  
For bellow-type gas flow meters, this is 
carried out on the input side of the meter 

 2 

:

 

For the MU: To connect the gas sensor 
of the Volume Conversion Device 

 3 

:

 

Test connection: Possibility to take test 
pressure or to gibe external pressure on 
the pressure sensor of the Volume 
Conversion Device. 

 

When installing the three-way cock, it must be noted that the position of the 
operating lever is checked with the corresponding throughputs, as the lever 
can be removed and may be installed in a twisted manner. 

Prüfen mit

Betriebsdruck

B

e

trie

b

s-

ste

llu

n

g

P

fe

n

 m

it

F

rm

d

d

ru

ck

Absperrung

45° Stellung

Содержание EK205

Страница 1: ...Operating manual Volume conversion device EK205 ...

Страница 2: ...Volume Conversion Device EK205 Operating instructions 73023159 Version d Issue date 16 12 2016 Software version from 1 02 ...

Страница 3: ...hazards 15 2 6 Environmental protection 16 2 7 Operator liability 16 3 Technical data 17 3 1 General data 17 3 2 Power supply for EK205 without integrated mains adapter 17 3 2 1 Battery supply 17 3 2 2 External power supply for the basic device 17 3 3 Pressure sensor 18 3 3 1 CT30 Type Pressure Sensor 18 3 4 Temperature sensor 18 3 5 Digital inputs 19 3 5 1 Pulse and reporting inputs NF 19 3 6 Dig...

Страница 4: ...s 29 5 3 2 Sealing 36 5 3 3 Close housing 37 5 3 4 Check installation and connection 37 5 3 5 Data transfer 37 6 Operation 38 6 1 Safety 38 6 1 1 Personal protective equipment 38 6 2 Operating staff 38 6 2 1 Trained individual 38 6 2 2 Specialist 38 6 2 3 Calibration officer 38 6 3 Basics 39 6 3 1 Display 39 6 3 2 Functions of the keys 40 6 3 3 Calling up the data navigation in the display 41 6 3 ...

Страница 5: ...taff 57 8 1 2 Personal protective equipment 57 8 1 3 Improper fault rectification 58 8 1 4 Conduct in the event of faults 58 8 2 Fault and other status messages 58 9 Appendix 62 9 1 Replacement and accessory parts list 62 9 1 1 Fastening parts 62 9 1 2 Pressure connection 62 9 1 3 Temperature sensor bags 63 9 1 4 Small parts and other 63 9 1 5 Documentation 63 9 2 Conformity declarations 64 9 3 De...

Страница 6: ......

Страница 7: ...le and may differ from the actual design of the device The data and material properties that are presented below are guidelines They must be reviewed for each individual case and corrected if necessary For the commissioning of the various communication and device applications you can use the application manual of the EK205 at www elster instromet com en 1 2 Warranty provisions You will find our cu...

Страница 8: ... not avoided WARNING indicates an immediately hazardous situation that may lead to death or severe injury if it is not avoided CAUTION indicates an immediately hazardous situation that may lead to minor or slight injury if it is not avoided indicates hazards from electric current In the event of non observance of the safety instructions there is a risk of severe or fatal injuries CAUTION indicates...

Страница 9: ...turer delivery conditions and current legal regulations that apply at the time the contract was concluded Read through these operating instructions carefully before beginning any work to and with the device especially before commissioning the device The manufacturer assumes no liability for loss and malfunctions that result from non compliance with these instructions We reserve the right to make t...

Страница 10: ... The replacement parts and accessories list can be found in the appendix Replacement parts and accessories can be ordered from a contract dealership or directly from our customer service The accessories of the EK205 also include the free enSuite program available at www elster instromet com With them you can program the Volume Conversion Device EK205 via its data interfaces in order to carry out f...

Страница 11: ...perature This may result in the device malfunctioning at a later time Therefore After the device has been stored or transported in cold weather or if it has been subject to extreme variations in temperature bring it slowly to room temperature before it is commissioned The device must undergo a waiting period of at least 12 hours before it is put into operation if condensation formed during storage...

Страница 12: ...pment only in accordance with the valid rules of the art stated in the design test certificates Connect the EK205 only to intrinsically safe pieces of equipment whose electric data correspond to the requirements stated in the design test certificates of the EK205 see chapter 9 3 DANGER Risk of explosion from use of wrong batteries Use only batteries stated in the design test certificates of the EK...

Страница 13: ...e as described here The Volume Conversion Device EK205 serves to convert a gas volume of a gas line in basic condition measured by a meter in measurement conditions as well as to assign the volumes measured to tariffs In addition using the device further parameters can be measured recorded and monitored depending on the configuration set by the user Compliance with all the specifications in these ...

Страница 14: ...or the specific location in which he or she works and is acquainted with the relevant standards and regulations Calibration officer has the ability because of his or her technical training knowledge and experience as well as his or her knowledge of the relevant standards and regulations to carry out calibration protected work to gas systems The calibration officer is trained to work with calibrati...

Страница 15: ...hapters to reduce health hazards and avoid dangerous situations WARNING Risk of injury when handling batteries incorrectly Batteries need to be treated with special care Therefore Do not throw batteries in the fire or expose them to high temperatures There is a risk of explosion Do not charge batteries There is a risk of explosion Liquid being released in the event of incorrect use may result in s...

Страница 16: ...ed The operator must keep himself herself informed of the applicable occupational health and safety regulations and determine over the course of a risk assessment the additional risks that arise from the specific working conditions when the device is being used The operator must include these items in the form of operating instructions for the device The operator must review over the entire operat...

Страница 17: ...n Value Unit Voltage 3 6 V General nominal capacity 16 5 Ah Usable capacity 13 0 Ah Minimally required number of batteries 1 Unit Minimal operating duration in standard operation mode 5 Years The standard operation mode is defined as follows Information Value Measurement cycle 30 seconds Mode input 1 Pulse input Display active 60 minutes per month Interface active 30 minutes per month Ambient temp...

Страница 18: ... abs 25 bar abs 2 0 10 bar abs 40 bar abs 1 0 16 bar abs 40 bar abs The pressure sensor is available as an internally completed variant You can find details at www elster instromet com or see chapter Installation connection and commissioning Not applicable when used as a Temperature Conversion Device 3 4 Temperature sensor Information Value Unit Measuring range 30 60 C Measurement uncertainty max ...

Страница 19: ...ain circumstances Information Value Unit No load voltage U0 2 0 V Internal resistance Ri About 500 kΩ Short circuit current Ik About 4 μA Switchpoint on Resistance Re max 300 kΩ Voltage Ue max 0 8 V Switchpoint off Resistance Ra min 5 MΩ Voltage Ua min 1 8 V Pulse duration te min 62 5 ms Break duration ta min 62 5 ms Meter frequency f max 10 Hz Cable length max 10 m 3 6 Digital outputs The digital...

Страница 20: ...ber of bus participants Drive performance at the outlet max 16 unit loads2 Received power at the input3 6 unit loads RS485 not electr insulated 3 unit loads RS485 electr separated Cable length max 10 m 3 9 Operating conditions 3 9 1 Environment Information Value Unit Temperature range 25 55 C Relative humidity maximum 93 Humidity conditions acc EN12405 1 condensation Point of use acc EN12405 1 ope...

Страница 21: ...6 Measurement accuracy data 7 Notices 8 Ambient temperature range 9 IECEx EC type examination certificate no 5 10 ATEX EC type examination certificate no 11 Serial number 12 Year of construction 13 Ex marking 14 Metrology marking 3 10 2 Device software identification Move the cursor with the arrow keys to the Serv tab and via the following path to the values Vers device software version or Chk che...

Страница 22: ...Cable bushing Temperature sensor 7 Outdoor plug option 8 Pressure sensor6 9 Optical interface 10 Security bails 4 2 Internal view Ill 3 1 Connections of the meter and reporting inputs DE1 and DE2 2 Connections of the pulse and reporting outputs DA1 and DA2 3 Connections of the serial interface 4 Connections of the battery 6 Not applicable when used as a Temperature Conversion Device 4 6 9 7 1 8 10...

Страница 23: ... temperature sensor When using the EK205 as a Temperature Conversion Device the momentary values of temperature are measured and the pressure is set as a constant As an operating element of the EK205 an alphanumerical display and a keyboard on the front side of the device are available 4 4 Connections The Volume Conversion Device EK205 has connection options for Two batteries To monitor and regist...

Страница 24: ...hten the self cutting screws such that the angle is tight 3 Using two M5 square nuts 4 and two M5 x 10 mm cylinder screws 2 install the mounting bracket on the rear of the meter head 5 4 Tighten the cylinder screws such that the device sits tight and cannot fall down 5 1 2 Installation on a pipeline Using pipe installation aids see appendix with suitable cable ties install the EK205 on a pipeline ...

Страница 25: ...the overall gas line The three way cock available from Elster has the following structure The pipeline from the pressure sensor to the meter must be laid in a falling manner so that where applicable water cannot damage the pressure sensor or influence the measurement precision 7 Not applicable when using EK205 as a Temperature Conversion Device Ill 7 1 From the meter From the pressure connection o...

Страница 26: ...Risk from faulty connection of the device The connection of the device may be carried out by a gas expert only see chapter Safety Faults in this process may result in fatal situations or considerable material damage Therefore Have the connection of the calibrated device be carried out by a gas expert only Also involve this person for subsequent relocations Do not perform any connection changes or ...

Страница 27: ... 1 in Ill 8 page 27 of the EK205 The priority can be freely selected The labels and of the terminals are intended for connecting other pulse transducers 2 Set the operation parameters such as the cp value pulse constant as described in chapter 5 3 1 2 5 2 2 Lead sealing of the input terminals After connecting to the gas meter in accordance with chapter 5 2 1 the input terminal DE1 needs to be seal...

Страница 28: ...sensor Pt 500 5 in the temperature sensor pocket 6 see appendix 2 Use the adapter for lead sealing the connection 2 see appendix 3 Fasten the temperature sensor through the capstan screw 3 and the planned screw connection 8 4 Have the temperature senor sealed using a lead sealing socket 1 and the wire seal 4 7 by a calibration officer 5 2 4 Connect pressure line8 When connecting the pressure lines...

Страница 29: ...tal outputs terminals DA1 and DA2 of the EK205 see Ill 12 2 For lead sealing the output terminal a corresponding terminal cover is provided in the accessories bag Where required screw them via the terminals and then glue a lead seal to the fastening screw see chapter 5 3 2 3 Where required for the pulse outputs set the cp value pulse constant as described in chapter 5 3 1 10 5 3 Commissioning 5 3 ...

Страница 30: ...s the ENTER button to confirm the set value It is possible to abort the entry by pressing the key combination ESC 2 Setting the cp value pulse constant Move the cursor to cp value cp I1 in the same path Press the key combination ENTER One digit of the value flashes Move the cursor with the arrow buttons or to the digits and change these with the arrow buttons or Press the ENTER button to confirm t...

Страница 31: ...pen the change information regarding setting Vm at the same value as the gas counter must be deleted Open the calibration lock Move the cursor to the Serv tab and via the following path to the value Clear adjustable volume at measurement conditions input mode Serv Volume Actual volume Synchronization Vm Clear Press the key combination ENTER 0 flashes Change the value using the arrow buttons or to ...

Страница 32: ...ditions input mode Serv Volume Standard volume Synchronization Vb Clear Press the key combination ENTER 0 flashes Change the value using the arrow buttons or to 1 After the value has been changed press the key combination ENTER in order to confirm the input It is possible to abort the entry by pressing the key combination ESC TimeX time stamp of the alignment Vb_o Vb old value Vb_n Vb new value ar...

Страница 33: ...ssible to abort the entry by pressing the key combination ESC 5 3 1 9 Setting replacement values for gas pressure and temperature If the gas pressure or the gas temperature cannot be measured due to a fault or if pressure and temperature are outside the alarm limits defined in chapter 5 3 1 8 an adjustable replacement value fixed value is used to calculate the standard volume The volume thereby ca...

Страница 34: ...le to abort the entry by pressing the key combination ESC Except for the settings described here many additional functions such as time synchronous pulses can be set for the outputs The complete description can be found in the application manual EK205 which you can download from www elster instromet com 5 3 1 11 Setting summer time switching Move the cursor to the Serv tab and via the following pa...

Страница 35: ...onfirm the input It is possible to abort the entry by pressing the key combination ESC 5 3 1 13 Deleting the certification data log Deleting the log is only possible when the calibration lock is open Move the cursor to the Serv tab and via the following path to the value ClCDL delete certification data log Serv Edit and delete ClCDL Press the key combination ENTER 0 flashes Change the value using ...

Страница 36: ...nts for securing the battery cover if available 5 Sealing points for securing the LCD plate cover The sealing of the sealing points Ill 15 3 for securing the pulse inputs and pulse outputs is subject to national law see WELMEC 11 1 chapter 2 7 1 Depending on the legal situation in the country of use seals from a group of people set out in law or from meter operators must be used If devices are del...

Страница 37: ...tion of the EK205 may result in fatal situations Therefore Carry out the installation and connection of the EK205 correctly Observe the requirements of the standards and guidelines stated in the design test certificates 5 3 5 Data transfer For the various options of remote data transmission to a control room via a device connected to the terminal interface please observe the corresponding instruct...

Страница 38: ...er behavior is authorized to read and note values and parameters using operating elements of the EK205 6 2 2 Specialist The specialist on operating the EK205 has the ability because of their technical training knowledge and experience as well as their knowledge of the relevant regulations to carry out the work to the device assigned to them and to recognize and avoid possible hazards on their own ...

Страница 39: ...esive label This needs to be operated to change calibration protected values and parameters In the event of a calibration protected device this may be carried out by a calibration officer only 6 3 1 Display The display is divided ex works into the five tabs of Main Cust Admin Serv and Ctrl in which the measurement values settings and other data are displayed Ill 19 Structure of the display 1 Data ...

Страница 40: ...lass and thereby activate the input mode open the relevant submenu update the relevant measurement value By pressing the two keys at the same time you can execute the ESC function regardless of the relevant data class and thereby leave a submenu and jump to its start value one menu level higher cancel an input the input mode By pressing the two keys at the same time you can execute the HOME CLR fu...

Страница 41: ...the other tabs to display their content The Customer tab contains more data than can be displayed simultaneously With the arrow key you move the cursor from the tab to the data field the bottom part of the display If the cursor is at the lowest visible value the display is moved by pressing the key again which makes the data visible Accordingly the cursor can be moved up with the key and the displ...

Страница 42: ... empty L12 Certification data log the certification data log is full P Calibration lock programming mode the calibration lock is open o online data transmission is ongoing If one of these letters flashes the condition thereby indicated is still in place active If a letter is displayed consistently the indicated status is no longer in place but needs to be confirmed Detailed explanations on the mea...

Страница 43: ...be opened to change this value 6 Invalid value The value entered is outside the permissible limits 7 Wrong key The key entered number code is incorrect lock is not opened 11 Input not possible due to special settings or configuration e g Change heat value only possible in the gas analysis list not in the energy list 12 The input of this source address is not permitted 13 Input not possible due to ...

Страница 44: ... volume conversion The calibration lock is designed as a key that is placed within the EK205 housing below the circuit board cover It can be secured with an adhesive label see chapter 5 3 1 1 Opening the calibration lock The calibration lock is opened by actuating the key the P symbol flashes in the display and closed again by actuating it a second time P symbol extinguishes Closing is also possib...

Страница 45: ...in encoder mode For every change of a parameter under CDL with a closed calibration lock a data row for the value is entered before and after the change If the certification data log is full it can be deleted when the calibration lock is open using the ClCDL command see chapter 5 3 1 13 If the calibration lock is opened while the calibration logbook is open it can only be closed again after deleti...

Страница 46: ...Display Meaning Unit Access Address Vb Volume at base conditions m3 C 2 300 VmT Total Volume at measurement conditions m3 C 4 302 p Pressure bar 7 310_1 T Temperature C 6 310_1 6 4 2 1 Vb volume in base mode The measured volume of the gas meter connected is converted in base mode and added up in this counter If a fault is given see chapter 8 Faults the counter is stopped and the volume is added up...

Страница 47: ...stomer tab This tab is used to display and check special device settings or statuses The application is intended for the gas customer This tab can be freely programmed by the user via the enSuite parameterization software The following parameters are programmed ex works Display Meaning Unit Access Address Time Date and time 1 400 Vb Volume at base conditions m3 E 2 300 VmT Total Volume at measurem...

Страница 48: ... what occurred since the last station inspection for instance The reports can be deleted on the device Serv Status Clr In the status tabs only alarms and warnings see chapter 8 2 are displayed 6 4 3 6 VbMP maximum measurement period counter Vb in the current month The time stamp of the maximum is displayed in the following row when moving the cursor to the VbMP value The maximum of the past 24 mon...

Страница 49: ... values17 Submenu for user specific parameters Volume Submenu for volumes and associated parameters Volume conversion Submenu for conversion and associated parameters Measured values Submenu for measurement values and associated parameters Archives Submenu for archives located in the device Status Submenu for current status status tab and logbooks Date and time Submenu for date and time and associ...

Страница 50: ...all outputs and associated parameters Interfaces Submenu for the interfaces in the device Device settings Submenu for general device settings Identification Submenu for the identification of the gas system Edit and delete Submenu for triggering various deletion functions Examination Submenu for freezing function and test archive Data book Submenu with information from the data book of the device 6...

Страница 51: ...k is closed open The calibration lock is open In this case the P at the top left also flashes in the display see chapter 6 3 4 Meaning of the status symbols 6 4 6 4 Menu selection of the display menu In delivery status the display of the EK205 has the five tabs Main Cust Admin Serv and Ctrl With the Menu value for certain purposes tabs can be displayed or hidden 6 4 6 5 Main content of the Main ta...

Страница 52: ...tivation Do not bridge any fuses or deactivate them When replacing fuses observe the correct ampere figure Keep life parts away from humidity This may result in a short circuit DANGER Risk of explosion due to electrostatic charging Avoid electrostatic charges Therefor Clean the device only with a damp cloth Avoid rubbing with non conductive materials WARNING Risk of injury when handling batteries ...

Страница 53: ...ult in severe personal injury or material damage Therefore Before starting work ensure sufficient installation freedom Ensure order and cleanliness at the place of installation Components and tools lying around and placed on top of one another loosely are sources of accidents If components were removed ensure correct installation reinstall all fastening elements and observe screw torques 7 1 2 Per...

Страница 54: ...es There is a risk of explosion Do not charge batteries There is a risk of explosion Liquid being released in the event of incorrect use may result in skin inflammations Avoid contact with the liquid Rinse the liquid with lots of water in the event of contact If the liquid enters the eyes rinse the eyes for 10 mins with water and see a doctor immediately The replacement of the batteries can be car...

Страница 55: ...scribed in chapter 7 2 3 In standard operation mode see chapter 3 2 1 the operation duration must display at least 62 months Otherwise compared the settings with the standard operation case and execute step 9 if applicable Ensure that the new battery is fastened correctly and sits tight inside the EK205 CAUTION Damage to property caused by improper device closure If the device is not properly clos...

Страница 56: ... settings Access St AL Press the key combination ENTER open flashes Move the cursor with the arrow keys or to change the value to closed After the value has been entered press the key combination ENTER in order to confirm the input It is possible to abort the entry by pressing the key combination ESC 7 2 3 Display remaining operating duration of the battery The calculation of the remaining operati...

Страница 57: ...f the external power supply carried out by electrical experts only For all work on the electrical plant de energize it and check the freedom from voltage Before maintenance work cut the voltage supply and secure it from reactivation Keep life parts away from humidity This may result in a short circuit 8 1 1 Staff The work described here for fault rectification unless indicated otherwise can be car...

Страница 58: ...ng on the type of the fault contact the manufacturer and have it rectified by authorized specialist staff or do so yourself 8 2 Fault and other status messages Faults here equivalent to alarms during the operation of the EK205 can be identified on the basis of the status symbols in the first row of the display see chapter 6 3 4 Further information and reports can be obtained in the Stat current st...

Страница 59: ...e see chapter 1 3 Customer service p Inp error The gas pressure cannot be measured due to a fault Please contact the Elster customer service see chapter 1 3 Customer service b Warnings Data restore The batteries of the EK205 have failed temporarily As a result of this the time has not changed and no measurements have been logged However all data remain preserved This message appears when old batte...

Страница 60: ...s Please contact the Elster customer service see chapter 1 3 Customer service Sett error The programing of the device has generated an unusable combination of settings Please contact the Elster customer service see chapter 1 3 Customer service I2 Warn sig If an input has been configured as a warning input this message will be displayed if an active signal appears at the corresponding terminal e g ...

Страница 61: ...ease contact the Elster customer service see chapter 1 3 Customer service Cal lock o The calibration lock is open In normal operating conditions the calibration lock should be closed so that unauthorized changes cannot be made To close the calibration lock please go to Ctrl St ES and enter the value 0 Admin lock Cust lock o The administrator lock customer lock is open Bat operat The EK205 is in ba...

Страница 62: ... 804 Mounting kit EK205 to S1 S1V S1D Enc S1 73 023 805 Universal holding angle for pipe installation for various nominal widths 73 021 955 Universal holding angle for fastening with clamp to a pipe 73 018 057 EK205 wall installation help 73 023 360 Pipe installation help incl screws 04 300 365 When ordering the universal holding angle with the clamps for pipe installation state the pipe diameter ...

Страница 63: ...EBL45 50 and 67 for fastening and sealing the 6 mm universal sensor 73 019 951 Adapter for older sensor bags EBL160 and 250 for fastening and sealing the 6 mm universal sensor 73 019 950 9 1 4 Small parts and other Replacement spare parts Order no Seal sleeve for pressure connection and temperature sensor 73 017 997 Connection cable for 2 NF pulse transducer approx 700 mm in length 73 017 093 Devi...

Страница 64: ...64 Appendix 9 2 Conformity declarations19 19 IECEx is not available for all device types ...

Страница 65: ...Appendix 65 9 3 Design test certifications 9 3 1 ATEX EC Type Examination Certificat ...

Страница 66: ...66 Appendix ...

Страница 67: ...Appendix 67 ...

Страница 68: ...68 Appendix 9 3 2 IECEx Certificate of Conformity20 20 Not available for all device types ...

Страница 69: ...Appendix 69 ...

Страница 70: ...70 Appendix ...

Страница 71: ...Appendix 71 ...

Отзывы: