Installation
, connection and commissioning
5.2.3.1 Connection to standard temperature sensor pocket
Ill.: 9
1.
Insert the temperature sensor Pt 500
4
in the
temperature sensor pocket
5
(see appendix).
2.
Fasten the temperature sensor through the
capstan screw
2
and the planned screw
connection
6
.
3.
Have the temperature sensor sealed using a
lead-sealing socket
1
and the wire seal
3
in
accordance with Ill.
9
by a calibration officer.
5.2.3.2 Connection to an older temperature sensor pocket
Ill.: 10
1.
Insert the temperature sensor Pt 500
5
in the
temperature sensor pocket
6
(see appendix).
2.
Use the adapter for lead-sealing the
connection
2
(see appendix).
3.
Fasten the temperature sensor through the
capstan screw
3
and the planned screw
connection
8
.
4.
Have the temperature senor sealed using a
lead-sealing socket
1
and the wire seal
4
,
7
by
a calibration officer.
5.2.4 Connect pressure line
8
When connecting the pressure lines, national requirements must be observed.
For Germany, the requirements of the PTB inspection rules, volume 20 Electronic
Volume Conversion Devices for gas, chapter 5 apply.
When laying pipelines, a falling laying must be observed.
8
Not applicable when using EK205 as a Temperature Conversion Device.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
Содержание EK205
Страница 1: ...Operating manual Volume conversion device EK205 ...
Страница 6: ......
Страница 64: ...64 Appendix 9 2 Conformity declarations19 19 IECEx is not available for all device types ...
Страница 65: ...Appendix 65 9 3 Design test certifications 9 3 1 ATEX EC Type Examination Certificat ...
Страница 66: ...66 Appendix ...
Страница 67: ...Appendix 67 ...
Страница 68: ...68 Appendix 9 3 2 IECEx Certificate of Conformity20 20 Not available for all device types ...
Страница 69: ...Appendix 69 ...
Страница 70: ...70 Appendix ...
Страница 71: ...Appendix 71 ...