Care and maintenance tasks
© Elma Schmidbauer GmbH
BA_Elmasolvex_SE_EN_Vers. 03.2018
37
8.3
Continuous actions
8.3.1
Change of cleaning and rinsing media
After view (visual inspection of the cleaning and rinsing media
in the opened media tanks) and/or in the event of diminishing
cleaning result.
If the media in the media tank seem to be increasingly
contaminated or the watch parts after cleaning no longer
appear to be clean, the media must be changed.
Only permitted media (cleaning / rinsing media) are permitted to
be used for safety reasons and to prevent machine damage.
Observe the instructions for recommended media and
limitations for unsuitable / not permissible media (
Chapter 8
)
.
Fire and explosion hazard!
Observe the applicable safety regulations for handling
solvents.
Keep all kinds of ignition sources away.
Prevent ignition sparks from electrostatic discharge. Discharge
possible electrostatic charges (body charge) before you handle
flammable media by touching any grounded equipment: e.g.
water tap, metal surface of the case of the cleaning machine or
use ESD protection equipment (ESD armband).
Remove the relevant media tank from the cleaning machine.
The media tank must not be removed during running cleaning
programs.
Drain the relevant media tank and clean if necessary.
Also examine the media tank for possible damage / glass
breakage and the seal for correct seating.
Used media must be disposed of in accordance with
the regulations. No disposal via the sewer system!
Dispose of the used media in accordance with the
national disposal regulations for the media (see safety
data sheet).
Recommended
interval
Inspection criteria
Only use permitted
operating materials.
DANGER
Procedure
ATTENTION
Checking media tank
Disposal of used
media