Care and maintenance tasks
38
BA_Elmasolvex_SE_EN_Vers. 03.2018
© Elma Schmidbauer GmbH
8.3.2
Removal of passed over materials
As required
Visually inspect the surfaces of the cleaning machine,
particularly the placement surfaces of the media tanks, for
entrained material residues.
In the case f larger media accumulations in a certain area,
check the media tank for glass breakage and the profile seal on
the media tank for damage.
Remove the residues with a dry, non-abrasive cleaning cloth.
8.3.3
Replacement of the filter mat
As required.
Reduced drying result and particle residues on cleaning parts
and/or on cleaning basket / movement holder after the drying.
Maintenance kit filter mat + fan grille Art. No. 104 9007
Only use the original seal of the manufacturer to prevent
impairment of the machine function.
Disconnect the mains plug first.
Undo the four Allen screws of the maintenance cover (
Fig.
8.3.3.1.A
) using an Allen key.
Remove the fan grille (
Fig. 8.3.3.2.B
) from the fastening at the
cable connector and remove it from the cleaning machine.
Remove the used filter mat (
Fig. 8.3.3.2.C
).
If required, clean the plastic fins on the cleaning machine using
a dry cloth.
Attach the new filter mat to the fan grille (the fan grille from the
maintenance kit is provided for reserve purposes in case the
fan grill has been damaged during removal).
Secure the maintenance cover with the 4 Allen screws.
Fig. 8.3.3.1. Opening service cover
Recommended interval
Inspection criteria
Action
Recommended interval
Inspection criteria
Order number
Procedure
A