Section 5: Setup
5.2 Check free movement of steel needle
1. Check initiation needle for free movement.
Open the Adjust Needle menu (
Start menu
Settings
Lab Options
Maintenance
Adjust
Needle
).
CAUTION!
Risk of injury from initiation needle!
The tip of the initiation needle is exposed. Movement of the initiation needle can cause needle puncture
injuries.
•
Always keep your hands and fingers clear from the area underneath the initiation needle.
WARNING!
Risk of infection due to sample residue and in case of inadequate hygiene, disinfection, and
sterilization procedures!
Exposure to sample residue in non-cleaned, non-sterilized, or non-disinfected components results in an
elevated risk of infection.
• Wear lab gloves during any kind of work.
• Observe all local regulations regarding hygiene, disinfectant cleaning, and sterilization.
2. Tap
Move Needle
and ensure that the initiation needle moves freely.
Tap
Back
to complete the check.
Figure 24 Adjust Needle Menu