32
Section 2: Design and Function
Result line
After each measurement a result line is sent which contains the following semicolon-separated values in a
fixed order:
Column
Description
batch
Batch identifier in quotation marks, entered by the user or a sequential number generated
automatically.
Or
Void for single measurements and if batch ID is disabled in the options.
sample
Sample identifier in quotation marks, entered by the user or a sequential number gener
-
ated automatically.
Or
Void if sample ID is disabled in the options.
date
Date and time in combined ISO 8601 format (e.g.:
2015-12-31T13:45
).
value
Measurement value as integer.
Or
Error identifier (see
Error messages
section).
dimension
Unit of the measurement value returned in
value
enclosed in quotation marks (
“mOsmol/
kg”
) – regardless of the selected result unit and the language setting of the device!
Or
Void if “
value
” contains an error message.
device-no
Serial number of the device in quotation marks.
check
Checksum of previous values in this line (see “Checksums” on page 35).
message
Human-readable message always in English in quotation marks.
Or
Void if there is no notification.
XML format
The XML format is sent line by line but a single record will generally extend across several lines. Each
record is transferred as a ticket and multiple measurements for one batch are combined into one ticket.
Strictly speaking, records are allocated to tickets in the same way as they are published: Each record
corresponds to one ticket in the XML log.
There are two types of tickets:
Type
Description
SAMPLE
Contains exactly one result from a single measurement.
BATCH
Contains several results from a batch measurement.
Содержание FreezePoint 6000 Series
Страница 1: ...FreezePoint Freezing Point Osmometer USER S MANUAL MODEL 6000 6000P 6000S 6000SP...
Страница 2: ......