Section 7: Troubleshooting
13. Push guide tube onto initiation needle and pass needle through stainless
steel tube of mounting plate.
Align the thermistor probe
14. Tighten the 2 mounting screws only lightly.
The thermistor probe must be aligned using the adjustment tool.
15. Install adjustment tool (1) in opening of lower cooling system using
tweezers.
16. Remove measurement vessel from thermistor probe.
Figure 93 Install adjustment tool
17. Align thermistor probe. Proceed as follows:
Push thermistor probe
just beyond
adjustment tool.
18. Position thermistor probe on XY plane so that thermistor is perpendicular
to center of hole of adjustment tool (“Figure 95 Thermistor probe (A) Z axis
19. Tighten the 2 mounting screws on the aligned thermistor probe.
Figure 94 Align the thermistor probe
Figure 95 Thermistor probe (A) Z axis and (B) XY plane