57
Section 6: Operation
11. Move elevator up.
NOTE!
Moving the elevator up during the measurement will abort the
measurement.
Take care when removing the measurement vessel to clear the sample
probe to avoid damaging it.
12. Remove measurement vessel from thermistor probe.
Dispose of measurement vessel and sample according to local regulations.
Figure 52 Clean thermistor probe
13.
Clean thermistor probe using a soft, dry, lint free paper towel.
NOTE!
Risk of carryover!
• Failure to clean the thermistor probe immediately following measurement can result in carryover and
incorrect measurement results.
• Cleaning the probe with a lint free cloth wetted with DI water then wiping and drying the probe, or
the use of water blank, may help alleviate this issue. This procedure is recommended to avoid risk of
contamination with salt at the thermistor.
14.
(If doing series or batch measurements) Start the measurement of the new sample by pressing “Next
Sample”, and repeating the work starting with step 9.
To complete the series measurement, tap Complete batch.
• On Model 6000P/Model 6000SP, completing the series measurment also completes the printing
operation.
NOTE!
Exiting the measurement menu (by pressing Quit) will also quit the series measurement.