Section 6: Operation
7. Position measurement vessel on thermistor probe with cover facing front
(1).
NOTE!
The measurement vessel is securely attached to the thermistor probe
when it clicks into place.
Figure 48 Position measurement vessel
8.
Tap Use printer on the measurement menu to outputthe measurement
results to a printer. Make sure that the printer is ready (1).
(Only Model 6000P/Model 6000SP)
9.
To start the measurement-
A.
For a single sample, tap “Single Sample” (2).
If necessary, enter the sample ID using the virtual keyboard.
OR
B.
To start a series measurement, tap “Begin Batch” (3).
Enter the batch ID.
Figure 49 Measurement menu
NOTE!
The sample/batch ID can be predefined in the user preferences. In this
case, the system assigns the sample/batch ID automatically.
10. Move elevator down.
NOTE!
Moving the elevator up during the measurement will abort the
measurement.
Be mindful when lowering the sample that droplets are not formed
on the side of the measurement vessel. This can affect the sample
reading.
The sample measurement is performed automatically. Pay attention to the
information displayed on the touchscreen.
Figure 50 Move elevator down
The measurement result displays on the touchscreen.
On Model 6000P/Model 6000SP the result is printed.
( “10.2 Performance parameters” on page 100)
Figure 51 Measurement