4.7.1.
ALARMS
By clicking on the button , allows to access the Alarms page
4.7.1.1.
REMOTE SETTING
The alarm appears when there is problem with the communication board
REMOTE SETTING ALARM does not cause any stop.
4.7.1.2.
CURRENT LIMITATION
TPS works a control of the output current and this allows it to support for an indefinite time the
output short circuit. In case of loads that absorb a current superior than the nominal one TPS
works a limitation of the same current. In case of current limitation the output wave is no more
guaranteed and so it will show an harmonic distortion.
Not linear loads with an overload less than the nominal one but with a very high crest factor
current allow the current defence.
CURRENT ALARM does not cause any stop.
Item
Name
Description
1
Home button
Allows to come back to the Home page
2
Advanced command mode
(1)
Allows to access the Advanced command mode page
3
Alarms button
Allows to access the Alarm page