9 • Use in PC mode
●
25
9.1 Port menu
From this menu, you can close an already open port and then choose
a new port.
9.2 Settings menu
9.2.1 Sort
If the sorting parameters have already been defined, a floating window
opens and displays the same information as in Stand-alone mode.
9.2.2 Set Sorting parameters...
As for the Stand-alone mode.
9.2.3 Set Sorting parameters...
As for the Stand-alone mode.
9.2.4 Firmware update...
This menu allows the firmware update. Normally, calibration data
should not be deleted (by default). Data will only be deleted if it
is mandatory for the update. In that case, it will be specified in the
instructions accompanying the update.
Occasionally, after selecting the file to be loaded (LCR6.hex, possibly
with a revision number), the update procedure does not start. If this
happens, close the AU2019 program, insert a jumper on J16, restart
the device, and restart the AU2019 program to repeat the update. Do
not forget, at the end, to remove the jumper on J16 before restarting
the device.
9.2.5 Texts update...
This menu permits a new set of languages used by the different menus
to be loaded into the instrument. A specially formatted text file (LCR6
Texts v3.txt to date) must be used for this purpose (see the Appendix
to this manual).
It can be downloaded from the author’s updates on the project page:
https://www.elektormagazine.com/labs/remake-lcr-meter
9.2.6 Calibrations
As for the Stand-alone mode.