Using the Delivery Modes
162
Infusion Parameters: PCEA Mode
The following infusion parameters need to be set for a PCEA infusion:
Parameter
Description/Notes
Continuous Rate
The speed at which the fluid is infused. Continuous Rate values can
range from 0.1 to 25 mL/h, or be equal to zero (Bolus Only
infusion).
VTBI
The total amount of fluid to be infused. The remaining VTBI is
displayed on the screen as the infusion progresses. VTBI values can
range from 0.1 to 9999 mL.
Demand Bolus
The amount of fluid infused in a single bolus. Demand bolus values
can range from 0.1 to 30 mL, or be equal to zero (Continuous Only
infusion).
Bolus Lockout
The minimum time that must pass between the end of one bolus to
the start of the next patient bolus. After a bolus delivery ends, the
next patient bolus becomes available following the lockout time.
Dose Limit
The option to choose if patient boluses are limited by number or
volume. When choosing No Limits, the patient boluses are set to
the maximum allowed volume, according to the other parameters
defined for the infusion, including Lockout Time and Demand
Bolus.
Содержание Q Core Sapphire
Страница 1: ......
Страница 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Страница 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Страница 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Страница 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Страница 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Страница 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Страница 288: ...Sapphire Infusion Pump User Manual 287 ...
Страница 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Страница 293: ......
Страница 294: ......