> To b e g i n a n e w P C A i n f u s i o n w i t h a D r u g L i b r a r y :
1.
From the Indicators Bar, verify that the pump is in PCA mode.
2.
From the
Start Up
screen, select
New Infusion
.
3.
If a warning that the air detection is disabled (OFF) appears, ensure that a set with
an air-eliminating filter is used, and press
OK
.
4.
From the
Drug Name
screen:
•
Using the keypad, enter the drug name, then press
Find
and proceed to
.
•
When the required drug is not found in the Drug Library, press
Choose
General
key on the toolbar:
From the Attention screen, confirm Choosing General and press
OK
.
(
To begin a new PCA infusion without Drug Library
and continue programming from there.
5.
From the
Drug List
screen, select the row of the relevant drug.
For more information about changing delivery modes, refer to
The
Find
key can be used to display all available drugs without
entering any characters (letters, numbers or symbols) or filter drug
names according to the characters entered.
'Choose General' will bypass drug specific limits, the infusion will be
programmed outside of the safe Drug Library environment.
To display additional drugs press
Next
.
Содержание Q Core Sapphire
Страница 1: ......
Страница 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Страница 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Страница 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Страница 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Страница 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Страница 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Страница 288: ...Sapphire Infusion Pump User Manual 287 ...
Страница 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Страница 293: ......
Страница 294: ......