Sapphire Infusion Pump User Manual
277
Average Bolus Volume After Occlusion
The following table presents the average time to a downstream occlusion alarm, and
the bolus volume after occlusion, at a rate of 25 mL/h.
Alarms
Refer to full list of alarms in
.
Prime
Manual or automatic priming (600 mL/h, or from air in
line alarm 900 mL/h)
Sensors
•
Air in line sensor:
Detects both single and
accumulated bubbles sized 0.02-0.5 mL. The desired
size range of each option can be selected by the
technician.
•
Upstream/Downstream occlusion sensor
•
Door open sensor
•
Temperature sensor
Parameter
Pressure
0.4 bar
1.2 bar
Average bolus volume
following downstream
occlusion
0.135 mL
0.75 mL
*
* Under single fault condition
Average time to downstream
occlusion alarm
38 sec
3 min
Parameter
Description
Содержание Q Core Sapphire
Страница 1: ......
Страница 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Страница 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Страница 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Страница 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Страница 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Страница 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Страница 288: ...Sapphire Infusion Pump User Manual 287 ...
Страница 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Страница 293: ......
Страница 294: ......