Sapphire Infusion Pump User Manual
201
Occlusion
pressure
The minimum downstream pressure that
triggers an Occlusion alarm. Acceptable
ranges are 5.8 to 17.4 PSI, 0.4 to 1.2 BAR
or 300 to 900 mmHg. An alarm sounds
when the downstream pressure reaches
the set value ± the sensor sensitivity level.
Select
Occlusion
pressure
. Then, using the
keypad, enter the desired
value
OK
.
Pump unattended
The number of consecutive minutes of no
interaction with the pump after which a
Pump Unattended
alarm is triggered.
Options are
2
,
5
or
10
minutes.
Note:
A Pump Unattended alarm is not
triggered when the pump is in Delayed
start or Standby.
Select
Pump
unattended
. Then,
select
2 min, 5 min
or
10 min
.
Infusion near end
The number of minutes before completion
of an infusion at which an
Infusion Near
End
alarm is generated.
Select
Infusion near
end
. Then, select
1 min
,
3 min
,
5 min
,
10 min
or
Off
.
Option
Descriptions/Notes
To Modify Parameter
(from the Alarms
screen)
Содержание Q Core Sapphire
Страница 1: ......
Страница 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Страница 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Страница 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Страница 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Страница 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Страница 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Страница 288: ...Sapphire Infusion Pump User Manual 287 ...
Страница 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Страница 293: ......
Страница 294: ......