Sapphire Infusion Pump User Manual
251
Troubleshooting
The following table lists some common programming issues, and explains how to solve
them.
Problem
Probable Cause
Solution
Programming cannot be
completed. The
OK
function key is disabled,
and the parameter
range is in red font.
The parameter entered is
outside of the safety range
calculated by the pump.
Verify the prescription, and obtain
a new one if necessary. Enter
infusion parameters within the
permitted ranges.
The
Set Delay
option
does not appear on the
Start
screen.
The option is not enabled.
Enable the
Allow delayed start
Authorization level is required.
The
PreSet Programs
option does not appear
on the
Start
Up screen in
any mode.
The option is not enabled.
Enable the
PreProgram
setting
). High Authorization level
is required.
The
Repeat last infusion
option does not appear
on the Start Up screen in
any mode.
The option is not enabled.
Enable
Repeat Last infusion
setting
). High Authorization level
is required.
Loading dose cannot be
programmed in PCA or
PCEA mode.
The option is not enabled.
Enable the
Allow loading dose
or
level of High is required.
Pump becomes locked
whenever an infusion
starts.
The Auto Patient Lockout
feature is enabled.
Disable the
Auto P. Lockout
option
). High Authorization level
is required.
Содержание Q Core Sapphire
Страница 1: ......
Страница 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Страница 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Страница 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Страница 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Страница 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Страница 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Страница 288: ...Sapphire Infusion Pump User Manual 287 ...
Страница 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Страница 293: ......
Страница 294: ......