EMC Tables
Page
32
/
35
Conducted
dis-turbances
induced by RF
fields
IEC 61000-4-6
3 V
0.15 MHz – 80
MHz, 6 V in
ISM bands
be-tween 0.15
MHz and 80
MHz, 80 % AM
at 1 kHz
3 V/m, 80 MHz
– 2,7 GHz,
80 % AM at 1
kHz
See the RF
wireless
communication
equipment
table in
"Recommende
d minimum
separation
distances"
3 V
Portable and mobile RF
communications equipment should
be usedno closer to any part of the
UltraMint
, including cables, than
the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended minimum
separation distances
See the RF wireless communication
equipment table in "Recommended
minimum separation distances"
Radiated RF
EM fields
IEC 61000-4-3
3V/m
Proximity fields
from RF
wireless
communication
equipment
Complies
IEC 61000-4-3
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare locations where
medical equipment and/or systems are used. When they are used in close proximity to medical
equipment and/or systems, the medical equipment and/or systems’ basic safety and essential
performance may be affected. The
UltraMint
has been tested with the immunity test level in the
below table and meet the related requirements of IEC 60601-1-2:2014. The customer and/or user
should help keep a minimum distance between RF wireless communications equipments and the
UltraMint
as recommended below.
Test frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
power
(W)
Distance
(m)
Immunity
test
level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
18Hz
1.8
0.3
27
450
430-470
GMRS 460
FM
2
0.3
28