Cleaning, Disinfection and Sterilization
Page
23
/
35
Functional
Testing,
Maintenance:
Visual inspection for cleanliness of the instruments and
reassembling.
Functional
testing
according
to
instructions of use. If necessary, perform reprocessing
process again until the instrument is visibly clean.
Before packaging and autoclaving, make sure that the
components have been maintained according to the
manufacturer’s instruction.
Packaging:
Pack the instruments in an appropriate packaging
material for sterilization.
WARNING
z
Check the validity period of pouch given by the
manufacturer to determine the shelf life.
z
Use pouches which resist to a temperature up to
141
℃
and in accordance with EN ISO 11607.
Sterilization:
Sterilization of instruments by applying a fractionated
pre-vacuum steam sterilization process (according to EN
285/EN 13060/EN ISO 17665) under consideration of
the respective country requirements.
Minimum requirements: 5 min at 134 °C.
Maximum sterilization temperature: 137°C.
Drying time: at least 8min.
Flash sterilization is not allowed on lumen instruments!
WARNING
z
Use only approved autoclave devices according to
EN 13060 or EN 285.
z
Use a validated sterilization procedure according to
EN ISO 17665.
z
Respect
the
maintenance
procedure
of
the
autoclave device given by the manufacturer.
z
Use only this recommended sterilization procedure.
z
Control the efficiency (packaging integrity, no
humidity, color change of sterilization indicators,
physicochemical integrators, digital records of cycles
parameters).
z
The sterilization procedure must comply with EN ISO
17665.