![Eighteeth UltraMint Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/eighteeth/ultramint/ultramint_user-manual_2369384021.webp)
6
Cleaning, Disinfection and Sterilization
Page
21
/
35
WARNING
z
Only the components above can be autoclaved.
z
Before first use and after each use, sterilize the above components.
Reprocessing Instructions
Preparation at the
Point of Use:
Before cleaning, disconnect the handpiece and tips from
the main unit. Disassemble the light guide and lamp
beads from the handpiece. Refer to Chapter 4.4 and 4.5
of this manual for disassembly instructions. Remove
gross contaminations from the components with code
water (<40°C) immediately after use. Don’t use a fixating
detergent or hot water (>40°C) as this can cause the
fixation of residuals which may influence the result of the
reprocessing process.
Store the instruments in a humid surrounding.
WARNING
Do not submerge the components or wipe them with any
of the following functional water (acidic electrolyzed
water, strong alkaline solution, or ozone water), medical
agents (glutaral, etc.), or any other special types of water
or commercial cleaning liquids. Such liquids may result
in metal corrosion and adhesion of the residual medical
agents to the components.
Transportation:
Safe storage and transportation to the reprocessing area
to avoid any damage and contamination to the
environment.
Preparation for
Decontamination:
The devices must be reprocessed in a disassembled
state.
WARNING
Observe suitable personal protective measures.
Pre-Cleaning:
Do a manual pre-cleaning, until the components are
visually clean. Submerge the components in a cleaning
solution and flush the lumens with a water jet pistol with
cold tap water for at least 10 seconds. Clean the
surfaces with a soft bristle brush.
Cleaning:
Regarding cleaning/disinfection, rinsing and drying, it is
to
distinguish
between
manual
and
automated
reprocessing methods. Preference is to be given to
automated reprocessing methods, especially due to the
better standardizing potential and industrial safety.