Model: UltraMint
ULTRASONIC SCALER
USER MANUAL
Страница 1: ...Model UltraMint ULTRASONIC SCALER USER MANUAL...
Страница 2: ......
Страница 3: ...stallation and removal 12 4 5 Tip installation and removal 13 5 Product function and use 14 5 1 Operation panel instructions 14 5 2 Function mode and use 15 5 3 Operating the device 17 5 4 Advanced se...
Страница 4: ...1 Overview Page 4 35 1 Overview 1 1 Content 8 6 4 5 3 1 2 7 1 Main Unit 2 Power Adapter 3 Foot Pedal 4 Handpiece Support 5 Handpiece 6 Tip 7 External Water Supply Hose 8 Wrench...
Страница 5: ...iece support 1 pc 6604008 External water supply hose 1 pc 6641006 Handpiece 1 pc Store in the Instrument Box 6651005 Wrench 1 pc Store in the Instrument Box 6651010 Tips 6 pcs Store in the following b...
Страница 6: ...erial number Class II equipment Type B applied part Alternating current Dispose of in accordance with the WEEE directive Keep dry 134 C Can be autoclaved up to a maximum temperature of 134 Celsius 55...
Страница 7: ...instruction Page 7 35 Date of manufacture LOT Lot of manufacture Authorized Representative in the European Community Manufacturer s LOGO Follow instructions for use Washer disinfector for thermal disi...
Страница 8: ...e the device to direct or indirect heat sources The device must be operated and stored in a safe environment 3 The device requires special precautions with regard to electromagnetic compatibility EMC...
Страница 9: ...ase power key Quickly decrease increase power key Power indicator Periodontal mode Scaling mode Endodontic mode Water control knob Handpiece cord Handpiece Tip Power display window Power adapter 4 2 B...
Страница 10: ...he main unit WARNING The power adapter must be plugged into the main unit before being plugged into the power supply 4 3 3 Install external water supply hose Install the external water supply hose to...
Страница 11: ...he range of 0 01 0 5MPa Lock nut External water supply hose 1 Unscrew the lock nut counterclockwise 2 Put the lock nut on the external water supply hose 3 Connect one end of the external water supply...
Страница 12: ...Removal Installation 4 4 2 Lamp bead protective shield and light guide and lamp bead removal and installation WARNING Disassemble the light guide and lamp bead from the handpiece before Cleaning disi...
Страница 13: ...ise until it turns 90 degrees Handpiece Tip Wrench Rotate the wrench until it turns 90 degrees NOTE 1 Pay attention to the gap of the wrench Use this gap 4 5 2 Tip removal 1 Align the gap of the wrenc...
Страница 14: ...it select the tip marked with P at the tip when using it Press the key to select the endodontic mode the indicator on the corresponding button is lit select the tip marked with E at the tip when 5 1 2...
Страница 15: ...ct function and use Page 15 35 using it 5 2 Function mode and use Water supply function and use Connect the other end of the external water supply hose to the connector that can provide water for the...
Страница 16: ...5 Product function and use Page 16 35 External water supply hose Water flow adjustment Adjust the water flow through the water control knob on the handpiece cord Water control knob Handpiece cord...
Страница 17: ...move the dental calculus Keep the water flow smoothly to avoid tip overheat Never use excessive force or stay for too long when cleaning the teeth 5 3 6 Power adjustment Adjust the power depending on...
Страница 18: ...The tip must be tightened and there must be fine spray coming out from the tip when operating Refer to the user manual of tips for detailed operating instructions 7 Change a new one when the tip or ha...
Страница 19: ...ng countdown the device will automatically stop and exit the Cleaning mode During the cleaning process step on the foot pedal again or press the key to stop cleaning and exit the Cleaning mode 5 4 2 H...
Страница 20: ...The device should no longer be reused in case of signs of material degradation In case of damage the device should be reprocessed before sending back to the manufacturer for repair 6 2 General recomm...
Страница 21: ...water strong alkaline solution or ozone water medical agents glutaral etc or any other special types of water or commercial cleaning liquids Such liquids may result in metal corrosion and adhesion of...
Страница 22: ...rior to use WARNING z Use only approved washer disinfectors according to EN ISO 15883 maintain and calibrate it regularly z Follow instructions and observe concentrations given by the manufacturer see...
Страница 23: ...ISO 11607 Sterilization Sterilization of instruments by applying a fractionated pre vacuum steam sterilization process according to EN 285 EN 13060 EN ISO 17665 under consideration of the respective c...
Страница 24: ...that the handpiece is intact and not damaged before sterilization or use and do not apply any protective oil on the handpiece z The two O rings on the handpiece cord and the handpiece insertion point...
Страница 25: ...e surfaces with a cloth lightly moistened with Ethanol for Disinfection Ethanol 70 to 80 vol at least 2min repeat for 5 times NOTE z Do not use anything except Ethanol for Disinfection Ethanol 70 to 8...
Страница 26: ...Maintenance When the device is not in use the power switch should be turned off and the power plug should be unplugged When not in use for a long time it should be powered and watered once a month for...
Страница 27: ...re is water spray The tip is loose Tighten the tip properly Tail wire failure Contact your local distributor or our company Handpiece failure Contact your local distributor or our company The connecti...
Страница 28: ...board damaged Contact your local distributor or our company Water seepage at the connection between the handpiece and the handpiece cord Damaged waterproof O ring Replace waterproof O ring The root c...
Страница 29: ...r 3W 20W Power adapter fuse T0 5AL250V Main unit fuse T1 6AL250V Water pressure at inlet 0 1bar 5bar 0 01MPa 0 5MPa Electrical safety class Class Applied part B Ingress protecting rating Ordinary equi...
Страница 30: ...ing domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC61000 3 2 Class A Volta...
Страница 31: ...50 300 cycle Mains power quality should be that of a typical commercial or hospital environment If the user of devices require continued operation during power mains interruptions it is recommended th...
Страница 32: ...00 4 3 3V m Proximity fields from RF wireless communication equipment Complies IEC 61000 4 3 Recommended minimum separation distances Nowadays many RF wireless equipments have being used in various he...
Страница 33: ...0 2 0 3 9 5500 5785 WARNING 1 Use of accessories and cables other than those specified or provided by the manufacturer of UltraMint could result in increased electromagnetic emissions or decreased el...
Страница 34: ...al Synthetic materials electrical components and printed circuit boards are disposed as electrical scrap Please deal with them according to the local environmental protection laws and regulation Right...
Страница 35: ...trict Changzhou City 213000 Jiangsu China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com Web www eighteeth com Llins Service Consulting GmbH Tel 49 175 4870819 Add Obere Seegasse 34 2...