STP-iX455/iXL455 Series Instruction Manual
9-8
O
pe
ra
tio
n
DO NOT insert or remove a communication cable while a communication device and a STP
pump are turned ON the power.
Communication interface circuit may break down if surge voltage caused by such as potential
difference of communication interfaces or static electricity is applied to communication line.
Communication failure occurs frequently by broken communication interface circuit.
RS485 is available with communication interface circuit according to the environment, but it
gets failure easily. Check the waveform of the differential signal is normal with measuring
instruments, such as an oscilloscope.
D+
D-
D+
D-
a) Normal
b) Abnormal
(abnormal voltage amplitude)
Figure 9.7 Example of differential signal waveform
Содержание STP-iX455
Страница 20: ...STP iX455 iXL455 Series Instruction Manual Installation This page intentionally blank ...
Страница 56: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Страница 64: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Страница 114: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Страница 120: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Страница 124: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Страница 136: ...STP iX455 iXL455 Series Instruction Manual 13 12 Troubleshooting This page intentionally blank ...
Страница 138: ...STP iX455 iXL455 Series Instruction Manual This page intentionally blank ...
Страница 146: ...This page intentionally blank ...