![Edwards STP-301 Series Скачать руководство пользователя страница 73](http://html.mh-extra.com/html/edwards/stp-301-series/stp-301-series_instruction-manual_2366166073.webp)
STP-301/451 Series Instruction Manual
Operation
Power input terminals for
the air cooling unit
Terminals for
the air cooling unit
Not used
POWER ON state
signal output terminals
L(L1)
AC POWER
N(L2)
N.O OUT
POWER
N.O OUT
N.O OUT
ALARM OUT COM
N.C OUT
BRAKE
N.O OUT
N.O OUT
NORM OUT COM
N.C OUT
ACC.
N.O OUT
REMOTE
N.O OUT
ALARM state signal
output terminals
BRAKE state signal
output terminals
NORMAL OPERATION state
signal output terminals
ACCELERATION state
signal output terminals
REMOTE selection state
signal output terminals
N.O: Normal Open
N.C: Normal Close
COM: Common
Figure 8.1 I/O TB1 Remote Output Signal Terminal Block