![Edwards STP-301 Series Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/edwards/stp-301-series/stp-301-series_instruction-manual_2366166030.webp)
STP-301/451 Series Instruction Manual
4-8
Installation
4.3.2
STP Pump Installation Positions
The STP pump can be installed vertically, horizontally, and upside-down and slanted.
Vacuum
Equipment
Upside-down
Horizontal
Slanted
Vertical
Figure 4.3 STP Pump Installation Positions
When installing the STP pump in a horizontal or slanted position, it is recommended to
install it so that the direction of the outlet port is on a vertical or horizontal plane in the
direction of the gravity.
This makes it possible to reduce the load on the magnetic bearing and the heat
generated by the STP pump.
Out let Port
Outlet Port
Outlet Port
Out let Port
Direction of Gravity
Figure 4.4 Positions of the Outlet Port on the Horizontally or Slanted Installed STP Pump