SE-3&SE-300 Series Electrocardiograph User Manual Settings
- 50 -
10/5mm/mV
,
the default value is
10mm/mV
10/5mm/mV
means that the gain of limb leads is set to
10mm/mV
, while
the gain of chest leads is set to
5mm/mV
.
AGC
AGC
means auto gain control.
Choose from:
On
or
Off
,
the default value is
Off
Select
On
,
the gain can be automatically adjusted according to actual
signals.
Baseline
Adjustment
Choose from:
Horizontal
,
Auto
or
Off
,
the default value is
Horizontal
Select
Horizontal
, the baselines of the lead groups are adjusted
simultaneously, and the baselines of the leads in the same row are on the
same line.
Select
Auto
, the baselines of the lead groups are adjusted respectively.
Select
Off
, the baselines of the lead groups are adjusted equally in the
ECG reports.
Paper Marker
Paper Marker
is used to identify the start point of each page of the
recorder paper.
Choose from:
Yes
or
No
,
the default value is
No
Select
Yes
if the paper with black markers on the bottom is used, and the
device can identify the start point of each page of the recorder paper while
printing ECG reports.