SE-3&SE-300 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
- 67 -
CAUTION
2. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.
3. Clean and disinfect reusable electrodes after each use.
WARNING
Turn off the power before disinfection. The mains supply must be switched off if it is used.
1.
Switch off the main unit and disconnect it from the power cord.
2.
Wipe the exterior surface of the electrocardiograph, patient cable, and reusable electrodes
(suction bulbs of chest electrodes and the clamps of limb electrodes) using a soft cloth
dampened with the disinfectant solution.
3.
Wipe off the disinfectant solution with a dry cloth after disinfection if necessary.
4.
Dry the electrocardiograph, patient cable, and reusable electrodes for at least 30 minutes in a
ventilated and cool place.
12.4 Care and Maintenance
12.4.1 Recharge and Replacement of Battery
1)
Capacity Identification
The battery capacity can be identified according to the battery symbol in the top right corner
of the LCD screen.
->
->
->
Capacity is from full to empty.
2)
Recharge
SE-3/SE-300 series is equipped with the recharge control circuit together with the built-in
rechargeable lithium battery. When the unit is connected to the mains supply, the battery will
be recharged automatically. Then the battery recharging indicator lamp (
) and the mains
supply indicator lamp (
) will be lit at the same time. During the recharging course, the
symbol
flashes in the top right corner of the LCD screen. When the battery capacity is
full, the symbol
stops flashing, and the battery recharging indicator lamp (
) is
black. When the 3-channel electrocardiograph is switched off, the battery recharging
indicator lamp (
) is black if the battery is fully recharged.