SE-3&SE-300 Series Electrocardiograph User Manual Settings
- 49 -
sequence. The start time of a lead group is just the end time of the
previous lead group.
Select
Synchronous
, the lead group is printed one by one in a certain
sequence. All leads are printed with the same start time.
Auto
Record
Length
Choose from:
Short (2.5s)
,
Medium (5s)
and
Long (10s)
,
the default
value is
Short (2.5s)
When
Auto Record Length
is set to
Short (2.5s),
ECG waves of each
lead group will be printed for about 2.5 seconds.
When
RECORD LENGTH
is set to
Medium (5s),
ECG waves of each
lead group will be printed for about 5 seconds.
When
RECORD LENGTH
is set to
Long (10s)
ECG waves of each lead
group will be printed for about 10 seconds.
NOTE:
1.
Auto Record Length
is corresponding with
Record Device,
Sampling Mode
and
Record Sequence.
2. The device acquires ECG signals from patients undergoing
short-term resting test. The time range can be 2.5s, 5s or 10s.
Manual REC Style
Choose from
3 channels
,
6 channels
and
12 channels
,
the default value
is
12 channels
Select a style to print the ECG waves in the manual mode.
Rhythm
Record
Mode
Choose from:
Save Paper
or
Quickly
,
the default value is
Save Paper
Select
Save Paper
,
10s after
pressing the
PRINT/STOP
key
on the
main
screen,
an ECG report is printed in the RHYT mode.
Select
Quickly
,
pressing the
PRINT/STOP
key
on the main screen
to
begin printing an ECG report immediately in the RHYT mode.
Speed
Choose from:
5mm/s
,
6.25mm/s
,
10mm/s
,
12.5mm/s
,
25mm/s
or
50mm/s
,
the default value is
25mm/s
In the manual mode, select
5mm/s
,
6.25mm/s
,
10mm/s
,
12.5mm/s
,
25mm/s
or
50mm/s
.
Only
25mm/s
and
50mm/s
are available in the Auto, OFF-Auto, RHTY,
and OFF-RHTY modes.
Only
25mm/s
is available in the R-R analysis mode.
Gain
You can set the indicated height of 1mV ECG on the paper.
Choose from:
2.5mm/mV, 5mm/mV, 10mm/mV
,
20mm/mV
or