SE-3&SE-300 Series Electrocardiograph User Manual Settings
- 59 -
External Input
The external input socket is equipped in the electrocardiograph, through
which the electrocardiograph can receive signals from the external
equipment.
Choose from:
On
or
Off
,
the default value is
Off
Select
On
, the electrocardiograph will display the signals which it
receives from external input
port.
External Output
The external output socket is equipped in the electrocardiograph, through
which the electrocardiograph can send rhythm lead signals to the external
equipment.
Choose from:
Off,
Standard
or
Triggered
,
the default value is
Off
Select
Standard
, the electrocardiograph sends ECG signals of rhythm
lead 1.
Select
Triggered
, the electrocardiograph sends pulses with the height of
5V and the width of 45ms, based on the data of rhythm lead 1.
Language
Select the language displayed on the main screen and in the ECG reports.
System Password
Type a password that allows you to access the
System Setup
window.
Load Factory
Settings
Press to restore the factory settings.