SE-1201 Electrocardiograph User Manual System Setup
- 87 -
Item
Description
Auto DispStyle
Choose from:
3×4
,
3×4+1R
,
3×4+3R
,
6×2
,
6×2+1R
or
12×1
Select
3×4
to display ECG waves of 12 leads in 4 groups of 3.
Select
3×4+1R
to display ECG waves of 12 leads in 4 groups of 3 with
the ECG wave of one rhythm lead on the bottom.
Select
3×4+3R
to display ECG waves of 12 leads in 4 groups of 3 with
ECG waves of three rhythm leads on the bottom.
Select
6×2
to display ECG waves of 12 leads in 2 groups of 6.
Select
6×2+1R
to display ECG waves of 12 leads in 2 groups of 6 with
the ECG wave of one rhythm lead on the bottom.
Select
12×1
to display ECG waves of 12 leads on one screen
simultaneously.
Manual Style
Choose from:
3 channels
,
6 channels
or
12 channels
Select
3 channels
to display ECG waves of 3 leads.
Select
6 channels
to display ECG waves of 6 leads.
Select
12 channels
to display ECG waves of 12 leads.
Rhythm Style
Choose from:
Single Lead
or
Three Leads
Select
Single Lead
c to print 60s ECG waves of the appointed single
rhythm lead.
Select
Three Leads
to print 20s ECG waves of three appointed rhythm
leads.
Sampling Mode
Choose from:
Real-time Sample
,
Triggered Sample
or
Periodic Sample
Select
Real-time Sample
, 10s ECG data sampled after pressing the
PRINT/STOP
key will be printed out.
Select
Triggered Sample
, after pressing the
PRINT/STOP
key, if
Arrhythmia ECG data, including Asystole, Ventricular
Fibrillation/Ventricular Tachycardia,
5>PVCS>=3, Paired PVCS, Bigeminy,
Trigeminy,
R ON T, single PVC and
Missed Beat, is
detected during the
learning course, the printing will be triggered automatically.
NOTE:
ID and patient information will not be changed while carrying
out the periodic printing.
Duration
& Interval
In the auto mode, when
Sampling Mode
is set to
Periodic Sample
, if
Interval
is set to
2 min
,
Duration
is set to
24 min
, after pressing the
PRINT/STOP
key, the printing will be performed every two minutes and
come to 12 times.
Auto Extend
Record
Choose from:
On
or
Off
Select
On
, if arrhythmia is detected in the auto or off mode, a hint will
pop up to ask you whether to print an extra rhythm report after the 12-lead
ECG report.
Содержание SE-1201
Страница 1: ......
Страница 67: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 60 6 5 4 ECG Reports in the R R Analysis Mode a ...
Страница 68: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 61 b ...
Страница 70: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 63 6 5 5 ECG Reports Printed by the USB Printer ...
Страница 100: ...SE 1201 Electrocardiograph User Manual System Setup 93 210 295mm 215 280mm ...
Страница 138: ... 131 ...