SE-1201 Electrocardiograph User Manual System Setup
- 103 -
Item
Description
Delete After Trans.
Or Export
Choose from:
On
or
Off
Select
On
, the files will be automatically deleted from the
File Manager
screen after they are transmitted to the PC or exported to the U disk or SD
card.
Replace When
Memory Full
Choose from:
On
or
Off
Select
On
, if the stored files reaches 200, the files will replace the earlies
one automatically.
Select
Off
, if the stored files reaches 200, the following hint will be
displayed.
If you select
Yes
, the current file will replace the earliest file stored in the
electrocardiograph.
If you select
No
, the current file will not be saved.
SCP File
Compression
Choose from:
On
or
Off
Select
On
, the SCP file will be compressed.
After the SCP function is activated,
SCP File Compression
appears in the
File Setup
window. For details on activating the SCP function, please
contact the manufacturer or the local distributor.