SE-1201 Electrocardiograph User Manual System Setup
- 88 -
10.3 Filter Setup
Select
Filter
on the
System Setup
screen, and then press
Enter
to open the
Filter Setup
window.
Figure 10-3 Filter Setup Window
Item
Description
AC Filter
Choose from:
On
or
Off
NOTE:
AC frequency can be set to
50Hz
or
60Hz
on the
Advanced
Setup
screen according to local mains supply
specifications.
DFT Filter
DFT Filter greatly reduces the baseline fluctuations without affecting the
ECG signals. The purpose of this filter is to keep the ECG signals on the
baseline of the printout.
Choose from:
0.05Hz
,
0.15Hz
,
0.25Hz
,
0.32Hz
,
0.5Hz
or
0.67Hz
(The set value is the low limit of the frequency range.)
EMG Filter
EMG Filter suppresses disturbance caused by strong muscle tremor.
Choose from:
25Hz
,
35Hz,
45Hz
or
Off
The cutoff frequency can be set to
25Hz
,
35Hz
or
45Hz
.
Select
Off
to turn off the function.
Lowpass Filter
Lowpass Filter
restricts the bandwidth of input signals.
The cutoff frequency can be set to
150Hz
,
100Hz
or
75Hz
.
All the input signals whose frequency is higher than the set cutoff
frequency will be attenuated.
NOTE:
Only when
EMG Filter
is set to
Off
, can the setting of
Lowpass Filter
be effective.
Содержание SE-1201
Страница 1: ......
Страница 67: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 60 6 5 4 ECG Reports in the R R Analysis Mode a ...
Страница 68: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 61 b ...
Страница 70: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 63 6 5 5 ECG Reports Printed by the USB Printer ...
Страница 100: ...SE 1201 Electrocardiograph User Manual System Setup 93 210 295mm 215 280mm ...
Страница 138: ... 131 ...