SE-1201 Electrocardiograph User Manual Technical Specifications
- 123 -
A1.5 Performance Specifications
Recording
Recorder:
Thermal dot-matrix recorder
Printing Density
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25
mm/s)
Recorder Paper:
Folded thermal paper:
210 mm×140 mm, 216 mm×140 mm, 210 mm×295 mm,
215 mm×280mm
Effective Width:
210mm
Paper Speed:
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s
(
±3%
)
Accuracy of data:
±5% (x-axis), ±5%(y-axis)
HR Recognition
Technique: Peak-peak
detection
HR Range:
30 BPM ~300 BPM
Accuracy: ±1
BPM
ECG Unit
Leads: 12
standard
leads
Acquisition Mode:
simultaneously 12 leads
A/D Resolution:
24 bits
Time Constant:
≥
3.2s
(0, +20%)
Frequency Response:
0.05Hz ~ 150Hz (-3dB)
Gain:
2.5, 5, 10, 20, 10/5, AGC (mm/mV)
Input Impedance:
≥
50M
Ω
(10Hz)
Input Circuit Current:
≤
0.01
μ
A
Input Voltage Range
≤
±5 mVpp
Calibration Voltage:
1mV±2%
DC Offset Voltage:
±600mV
Noise:
≤
12.5
μ
Vp-p
Multichannel crosstalk
≤
0.5mm