SE-1201 Electrocardiograph User Manual Transmitting ECG Data
- 67 -
c)
The last section of the
Local IP
item can be set at random.
For the same network transmission,
a)
Set the first three sections of the
Local IP
item to the first three sections of the IP of
the PC.
b)
The last section of the
Local IP
item can be set at random, but it can’t be the same as
the last section of the IP of the PC.
4)
Set the
Gateway
item.
Set the first three sections of the
Gateway
item to the first three sections of the IP of the
electrocardiograph. The last section of the
Gateway
item must be set to
1
.
5)
Set the
Subnet Mask
item to
255.255.255.0
.
6)
Press
Enter
to confirm, and then press
Esc
to return to the main screen.
4.
Set
Auto Transmission
to
On
in the
Transmission Setup
window.
5.
In the auto or rhythm mode, ECG data will be transmitted through the net cable automatically
after an ECG report is printed out. In the off mode, the sampled ECG data can be saved and
will be transmitted through the net cable automatically after the
PRINT/STOP
key is pressed,
but it can not be printed.
Содержание SE-1201
Страница 1: ......
Страница 67: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 60 6 5 4 ECG Reports in the R R Analysis Mode a ...
Страница 68: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 61 b ...
Страница 70: ...SE 1201 Electrocardiograph User Manual Printing ECG Reports 63 6 5 5 ECG Reports Printed by the USB Printer ...
Страница 100: ...SE 1201 Electrocardiograph User Manual System Setup 93 210 295mm 215 280mm ...
Страница 138: ... 131 ...