6.3.3. Init/Stop/Reinit Program
For controlling initialization /stop/reinitialization of program.
T
EXT
O
PTIONS OF
M
ENU
I
TEM
Setting
Description
INIT PROGRAM
program initialization. Displays in menu after loading program from menu
item
LOAD PROGRAM
) after program termination.
STOP PROGRAM
Program stopping. Displays in menu only during initialization phase of a
program.
REINIT PROGRAM Returning program to initialization phase. Displays in menu all time during
running program.
Press ENTER to proceed selected action.
6.3.4. Program Setup
Here can be unloaded, viewed, created and eventually edited gradient programs and
program mode is abandoned. System enable to create and save 12 programs. Default status is
EMPTY
.
6.3.4.1. Load Program
By uploading a program into unit memory is the unit prepared for initialization by
INIT
PROGRAM
. System allows to upload only one program.
Use arrows to select desired program, press ENTER to upload the program and then is the
menu item automatically abandoned. Cursor on the screen points at menu item
INIT PROGRAM
.
Press key ENTER again to set initialization phase parameters.
6.3.4.2. View Program
Viewer of created programs enable viewing set parameters
LOAD PROGRAM
. In heading is
written
VIEW PROGRAM
.
View Time Table
Sequence number of step, time (min:s) of step start, flowrate (ml/min), mobile phase
concentration A, B, C, D in % of volume. Use arrows up and down to move between table pages.
An unadjusted valve is marked with
- -
.