Connect outlet capillary.
Tighten capillary nut with side
wrench 1/4"
Connect the other end of capillary
to injector valve.
4.3. Injector Holder and Column Holder Mounting
Use Torx screwdriver TX20 to
unscrew two screws in front part of
the right panel of the unit.
Use the unscrewed screws to
mount the injector holder,
according to the picture.
4.4. Injection Valve Mounting
Injection valve is not a part of pump accessories, neither a part of the entire system
accessories and shall be ordered separately.
The injection valve holder is constructed for mounting two different types of valves and with
hole for outlet capillary from pump to injection valve
– Rheodyne and Vici type. The difference is in
spacing of holes for mounting the valve, see picture below.
Spacing 25.4 mm is designed for injection valve
Rheodyne.
Spacing 27.5 mm is designed for injection valve
Vici.
Spacing 12.5 mm is designed for outlet capillary
from pump to injection valve.
All diameters are in millimeters (mm).
Injection valve Rheodyne
is
mounted to the holder with two
screws using allen/imbus key.
Usually it is among the injection
valve accessories.