ECP2050
52 / 64
6.7.1.4. Needle of Degassing/Purging Valve
Disconnect the unit from power
source!!!
Unscrew needle.
Screw new needle.
6.7.1.5. O-ring in Degassing/Purging Valve
Disconnect the unit from power
source!!!
Unscrew two screws using
screwdriver TX20 from unit
accessories.
Take out the degassing/purging
valve. Under head, there is white
or transparent seal.
Unscrew needle, take out metal
ring and O-ring.
Take in new O-ring.
After replacement of O-ring, steps for mounting are the same as for dismounting, only in
reverse order.
Caution:
Needle is screwed into head of degassing/purging valve so that it does not
go out and then screw head of valve with needle onto compensating block.
6.7.2. Calibrations
In these parts of menu is calibrated pressure sensor, flowrate and gradient. For user is
available only automatical calibration of pressure sensor zero
OFFSET AUTO
and automatical
flowrate calibration
FLOW AUTO
.
C
ALIBRATION
F
LOW
C
HART