10.4.3.1 Preparation
Fig. 80: Calibration pump with oval gear meter (OGM
PLUS
)
1.
Connect the pump ready for operation on the pressure side
Chapter 7 ‘Device installation’ on page 32
2.
Fill a suitable measuring cylinder and insert the suction line or suction and return
line.
NOTICE!
The volume of the measuring cylinder should be 1/50 of the metering pump
rate in litres/h. During this process, the suction line must not change its
position.
The calibration of the metering pump is only valid for the currently set stroke
length setting. After changing the stroke length, the calibration must be
performed again.
Menu description
102
417101442 Rev. 12-02.2019