10.4.2.2 Calibration / Start
Fig. 79: Kalibrierung /Starten
The pump runs 122 strokes (ca. 1 min).
The strokes are counted down from 122 to 0.
After the running process, the amount removed from the measuring cylinder
(
Chapter 10.4.2.1 ‘Preparation’ on page 100
) must be read off. This quantity is then
entered as the calibration value (in ml or l).
NOTICE!
By activating the 'Menu/Exit’ function, the running calibration can be
interrupted.
10.4.3 Calibration pump with oval gear meter (OGM
PLUS
)
If a dongle box and an oval gear meter are connected and the “oval gear meter” function
(
Chapter 10.3.11 ‘Configuration / Oval gear meter (only E60
) is activated, the oval gear meter pulse rate recorded is automatically
displayed at the end of the calibration process.
This is offset against the specified calibration value. A correlation thus occurs between the
oval gear meter pulses and the metered quantity.
Menu description
101
417101442 Rev. 12-02.2019
Содержание Elados EMP KKS E60
Страница 2: ......
Страница 123: ...14 2 Abmessungen Abb 89 Abmessungen Technische Daten 121 417101442 Rev 12 02 2019 ...
Страница 134: ...16 Konformitätserklärung Abb 95 Konformitätserklärung Konformitätserklärung 132 417101442 Rev 12 02 2019 ...
Страница 135: ......
Страница 258: ...14 2 Dimensions Fig 90 Dimensions Technical Specifications 122 417101442 Rev 12 02 2019 ...
Страница 269: ...16 Declaration of Conformity Fig 96 Declaration of Conformity Declaration of Conformity 133 417101442 Rev 12 02 2019 ...
Страница 393: ...14 2 Dimensions Fig 93 Dimensions Données techniques 123 417101442 Rev 12 02 2019 ...
Страница 404: ...16 Déclaration de conformité Fig 99 Déclaration de conformité Déclaration de conformité 134 417101442 Rev 12 02 2019 ...
Страница 405: ......