35
7.
Make sure both O-rings are on the threaded connector.
8.
Slide the PCB into the enclosure until the two tangs “click”.
9.
Put the nut on and tighten ONLY ½ turn from where it is seated. If this nut is
tightened too much, the connector will crack and the enclosure will not be sealed.
10.
The module is now in EIB mode and can be on the EIB with a display.
To convert a module from EIB to Stand-alone Mode
The process is similar to the previously-described procedure. Note that in EIB mode, the
module will not show up in the device list of the Configuration Tool.
1.
Use the Configuration Tool (software) to “Set to Stand-alone Mode”.
2.
Install the jumper on JP4 in the module.
3.
In Stand-alone Mode, the module will continuously broadcast data via 500 kHz CAN
(programmable). For more information, refer to the NH
3
CAN Module Instruction
Manual.