17
another module, this user calibration will go with the sensor and be used with the new
module.
SPAN
To perform a SPAN:
1.
A span should be performed after the NH
3
sensor and pressure sensor (if so equipped)
have been on for at least 20 minutes.
2.
Put the NH
3
sensor and the pressure sensor (if so equipped) in the model gas
calibration apparatus shown in Figure 6 or in the exhaust of an engine.
3.
For the case of the model gas calibration apparatus, adjust the flowrates to get
approximately 15% O
2
, 1.5% H
2
O, your target calibration NH
3
ppm (typically 100),
balance N
2
. For the NH
3
tank concentration shown and a target of 100 ppm NH
3
, the
flowrate through flowmeter #2 will be about 1/9
th
that through the bubbler. This will
give the approximate 1.5% H
2
O required. To set the %O
2
and NH
3
ppm level, use a
NOxCANt or NOx 5210T analyzer. With the Type T NOx sensor used in these
devices, NH
3
will be read as NOx. The NH
3
must not be bubbled nor should there be
any condensed water from the output of the bubbler to downstream of the NH
3
sensor.
Otherwise, the condensed water will absorb the NH
3
and the span process will be
problematic.
4.
For the case of using the exhaust of an engine to span the NH
3
sensor, you must know
the wet NH
3
concentration from another instrument.
5.
Press the SYS key until “MOd” appears.
6.
Press the
key until “CAL” is on the display of the channel to be calibrated. Then
press the ENT key.
7.
Press the
key until “NH3” appears. Then press the ENT key.
8.
Press the
key until “SPAN” appears. Then press the ENT key.
9.
Using the
and
keys, change the display to read the actual NH
3
ppm.
10.
Press SYS to return to RUN mode.
11.
The user calibration is written into the memory chip in the NH
3
sensor’s connector
and will be used to calculate NH
3
. If the NH
3
sensor is removed and installed on
another module, this user calibration will go with the sensor and be used with the new
module.
FACT (return to factory NH
3
calibration)
To return to the factory calibration for the NH
3
sensor:
1.
Make sure the NH
3
sensor is attached to the module.
2.
Press the SYS key until “MOd” appears.
3.
Press the
key until “CAL” is on the display of the channel to be calibrated.
Then press the ENT key.
4.
Press the
key until “NH3” appears. Then press the ENT key.
5.
Press the
key until “FACT” is on the display. Press the ENT key. The user
NH
3
ZERO and SPAN calibration of the NH
3
sensor is erased and the factory
calibration will be used to calculate NH
3
.