Instruction Manual MM850481 ed. 03a
8
Low Draft Probe
x
Range
:
r
200Pa
x
Resolution
: 0.1Pa
x
Operative temperature range
: from +5 to +45 °C
x
Accuracy
From +10 Pa to -10 Pa:
r
0.5 Pa
Elsewere:
r
3 Pa
1.3 Accuracies
and
ranges
Parameter
Sensor type
Range
Resol.
Max
response
Accuracy
O
2
Electrochemical
from 0 to 25.0%
0.1%
20 sec.
r
0.1% vol.
CO
H
2
compensated
Electrochemical
from 0 to 8000 ppm
1 ppm
50 sec.
r
10 ppm <300 ppm
r
4% rdg up to 2000 ppm
r
10% rdg elsewere
CO diluted
Electrochemical
from 0.8 to 10%
0.05%
CO
Electrochemical
from 0 to 20000 ppm
1 ppm
40 sec.
r
10 ppm <300 ppm
r
4% rdg up to 2000 ppm
r
10% rdg elsewere
CO
%
Electrochemical
From 0 to 10.00%
0.01%
50 sec.
r
0.01% up to 0.2%
r
4% rdg elsewere
NO
Electrochemical
from 0 to 4000 ppm
1 ppm
40 sec.
r
5 ppm <125 ppm
r
4% rdg elsewere
NO LOW
Electrochemical
From 0 to 500 ppm
0.1 ppm
40 sec.
r
2 ppm <40 ppm
r
5% rdg elsewere
NO
2
Electrochemical
from 0 to 1000 ppm
1 ppm
50 sec.
r
5 ppm <125 ppm
r
4% rdg elsewere
NO
2
LOW
Electrochemical
from 0 to 500 ppm
0.1 ppm
40 sec.
r
2 ppm <40 ppm
r
5% rdg elsewere
NO
x
Calculated
from 0 to 5000 ppm
1 ppm
SO
2
Electrochemical
from 0 to 4000 ppm
1 ppm
40 sec.
r
5 ppm <125 ppm
r
4% rdg elsewere
CO
2
Calculated
from 0 to 99.9%
0.1 %
Tair
Pt100
from -10 to 100 °C
0.1 °C
r
(0.2% rdg+0.15°C)
Tgas
Tc K
from -10 to 1000 °C
0.1 °C
r
(0.3% rdg+0.3°C)
'
T
Calculated
from 0 to 1000 °C
0.1 °C
Tflow
Tc K
from -10 to 100 °C
0.1 °C
r
(0.3%rdg+0.3°C)
Treturn
Tc K
from -10 to 100 °C
0.1 °C
r
(0.3% rdg+0.3°C)
Pressure
'
P
Piezo
from 0 to 200 hPa
0.1 hPa
r
0.5 Pa up to 50 Pa
±1% rdg elsewere
Draft
Piezo
±40 Pa
0.01 hPa
r
0.03 Pa up to 300 Pa
±1% rdg elsewere
Excess air
Calculated
from 1.00 to infinity
0.01
Efficiency
Calculated
from 0 to 99.9%
(also for condensating
boilers with automatic
detection)
0.1 %
Gas speed
Pitot tube
From 0 to 100.0 m/s
0.1 m/s
Smoke index
External pump
from 0 to 9
Technical units and ranges can be converted directly from ppm to mg/Nm
3
, mg/kwh and from hPa to mmH
2
O, mbar or inH
2
O.
The relative accuracy shown are expressed as absolute or % of rdg errors at -5°C to +40°C ambient temperature.
The maximum response time shown is referred to 90% signal changes.
The pressure relative accuracy shown is valid only after the auto-zero procedure.
Specifications may change without notice.