Instruction Manual MM850481 ed. 03a
30
x
Press
[
ENTER
] to manually change the
'
P value.
The instrument will display the gas speed.
Press [
ESCAPE
] to the previous page menu.
NOTE: F
OR THE BEST SPEED CALCULATION
,
DO NOT CHANGE THE BURNER RATE BETWEEN THE TWO MEASUREMENTS
. T
HE
O
PTIMUM
CONDITIONS WILL BE ACHIEVED USING
2
HOTELS IN CHIMNEY AND THE SIMULTANEOUS MEASUREMENTS
7.8
Analysis data print
An internal impact printer can be installed as an option in the instrument. Read paragraph 8.10 for header report changing.
To print the report (ticket) of the analysis use the following procedure:
Before make all measurements you need: Smoke index, pressure and draught, T. flow, etc. After make the gas analysis
measurements and wait same seconds for measurement stabilizing. Press the [
] key. The following page will be
displayed:
ANALYSIS SAVED
PRESS [PRINT]
NUM. COPY 1
MODEL Standard
ESCAPE
–
FEED
Use
the
[
S
] and [
T
] keys to move the cursor symbol to “NO. copy” or “Model” field.
Press the [
ENTER
] key to change the number of the reports you have to print and/or the print model for the report.
Press the [
] key again to print the report. You can abort the printing in progress by pressing the [
ESCAPE
] key.
Press
[
ESCAPE
] (F1) to abort the operation.
Press
[
FEED
] for paper advance. You can stop the paper advance by pressing the [
ESCAPE
] key.
7.8.1
Paper and Ribbon change
Open the cover pressing as in figure.
Extract the exhausted paper roll. You can use standard paper roll but you have to check before for its dimensions (max.
57mm ±1mm large and 48mm ±1mm roll diameter).
If necessary substitute the ribbon cartridge.
Approach the end of the new paper roll to the slot under the printer.