Instruction Manual MM850481 ed. 03a
33
7.9.2
Test B (programmable)
This optional operative mode allows operator to find pipe leakage. The difference from Test A is the possibility for the use of
programmable the time, to see ( the next paragraph).To set and run leak test, press the [
MENU
] key from measuring page.
The following page will be displayed:
MENU
MEASURE
ION. CURR
TEMPERATURE
GAS VELOCITY
ESCAPE
–
Use
the
[
S
] and [
T
] keys to move the cursor symbol to “MEASURE” option.
Press the [
ENTER
] key to select the option. Press [
ESCAPE
] (F2) to return to the previous menu.
LEGEND
Time of the test
Gas Type: City or Natural
Barometric Pressure
P1: The Pressure initial
P2 : The Pressure end
System vol. Total Plant Volume
Counter vol. Gas Meter
Test[Air]
Diff of pr.
ǻ
P difference of pressure between P1 and P2
Dispersion found Capacity of Gas Dispersed
Dispersion correct Capacity of Gas correct
Suitability Conformity of Plant
ECLIPSE
EGA4
Serial Number 000
Date: 00/00/0000
Time: 00:00
P DECAY LEAK TEST
TEST A
Gas Type: City
Barom.Press.1013
mbar
P1: 0.0 mmH
2
O
P2: 0.0 mmH
2
O
System vol: 0.0 dm
3
Counter vol: 0.0 dm
3
Test[Air] 0.0 mmH
2
O
Diff of pr.0.0 mmH
2
O
Dispersion correct
NaN
dm
3
/h
Suitability: Yes(*)
PLANT
TAG
Plant 1
BOILER
TAG
Boiler 1
CUSTOMER
Name: Customer 1
Addr.:
City :
Te
l.: :
Note : -------------
--------------
--------------